Re: [DIYbio] Help Fix an Invitrogen Blue Light Transilluminator

Thanks for your help everyone!  I measured the current across the fuse while running at full brightness as 210mA.  I also noticed that next to the fuse was a spot to solder in a cartridge fuse holder and bypass the surface mount fuse.  I got a cartridge holder from Digikey (part# F6094-ND) as well as a 400mA fuse.  Soldered it in and removed the old surface mount fuse.
http://i.imgur.com/DAFgL3y.jpg
"F1" is where the old fuse was.

Everything works great now.
http://i.imgur.com/JkbM4Uh.jpg

Time to go stain some gels with SYBR Safe :)

-cory


--
-- You received this message because you are subscribed to the Google Groups DIYbio group. To post to this group, send email to diybio@googlegroups.com. To unsubscribe from this group, send email to diybio+unsubscribe@googlegroups.com. For more options, visit this group at https://groups.google.com/d/forum/diybio?hl=en